AK 98 Dialysis Machine - Disinfection and cleaning
A:151
WARNING!
After a chemical disinfection
program, a test for residues
shall be performed prior to
connecting a patient to the
dialysis machine to make
sure that no residues exist in
the dialysis machine.
10.4.4
Chemical disinfection program with a WRO unit
The dialysis machine and a WRO water purification unit can work together and do an
integrated chemical disinfection. Both the dialysis machine and the WRO unit must
be preset by an authorised service technician. For more information, see Section 11
“Disinfection with the AK 98 dialysis machine and WRO system”
on page A:159.
10.4.5
Test for disinfectant residues
After a chemical disinfection the dialysis machine shall be tested for disinfectant
residues. The fluid path must not contain chemical disinfectant residues, which can
cause harm to a patient.
Use an appropriate test method, either with proven sensitivity for the chemical or
recommended by the manufacturer of the chemical disinfectant. Limit values for levels
of disinfectant residues are specified by the clinic or by national standards.
Procedure
1)
Press the
On/Off
button to start the
dialysis machine.
When functional check is completed with
the proper concentrates attached, and the
green fluid path lights up:
2)
Press the
Fluid bypass
button.
3)
Collect a test sample of the dialysis fluid
from the fluid outlet tube going to the
dialyzer.
4)
If the test shows disinfectant residues:
Remove and discard the BiCart cartridge
and disconnect the Liquid A concentrate.
5)
Press the
Disinfection
button.
6)
Select the
Rinse
tab.
7)
Press
Rinse
.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...