AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
221 UF volume deviation
221 UF volume deviation
Second UF volume deviation {0} mL. Check
patient weight loss. Discontinue treatment and
call service.
Appears:
When the ultrafiltration volume measurements, for the
second time, are not within the specification of the dialysis
machine.
This can happen when:
●
the UF control is not calibrated or incorrectly calibrated.
●
the UF control does not work correctly.
●
the protective UF sensor does not work correctly.
Machine actions:
The dialysis fluid bypasses the dialyzer.
Operator action:
End the treatment and call an authorised service technician.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...