HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Attentions
B:85
510 Blood pump overloaded
510 Blood pump overloaded
Check blood pump for obstruction. To continue
press Confirm.
Appears:
When the blood pump is overloaded and can hardly run in
functional check.
Machine actions:
The blood pump stops.
Operator action:
1.
Check that the blood pump segment is correctly placed,
see Section 4.2.4
“Attach the arterial blood line”
on
page A:79 in Operators handbook.
2.
Press
Confirm
to continue.
If the attention appears again, contact an authorised
technician.
Mute time:
Permanent.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...