B:158
AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
588 Set fluid flow is not reached
588 Set fluid flow is not reached
Adjust dialysis fluid flow or call technical
service.
Appears:
The actual dialysis fluid flow rate is not the same as the set
value. The dialysis machine cannot set the required dialysis
fluid flow rate.
Machine actions:
None.
Operator action:
If the treatment can continue with a lower flow, adjust the
set value to the actual value.
If this is not possible, contact an authorised technician.
Mute time:
Permanent.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...