A:148
AK 98 Dialysis Machine - Disinfection and cleaning
HCEN12745 Revision 10.2016
Program version 2.xx
2)
Select the
Tools
tab.
3)
Press
Auto schedule
.
4)
Press
Rinse/Disinfection process
.
5)
Select
Heat
.
6)
Press
OK
.
7)
Select day.
8)
Set the time when the program should
finish.
9)
Press
OK
.
10)
To schedule more days in the week: repeat
steps 7-9.
11)
To perform a heat disinfection with
CleanCart cartridge: Place a CleanCart
cartridge into the cartridge holder before
the disinfection starts.
10.3.7.3
To turn off a scheduled program
Procedure
1)
Press the
Disinfection
button.
2)
Select the
Tools
tab.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...