HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
A:79
7)
Close down the upper latch to secure the
cartridge. Hold with one hand underneath
the lower latch.
8)
Attach the dialyzer (1) in the holder.
If necessary, squeeze the spring-clip (2) to
place the dialyzer in the holder.
9)
Hang the saline or priming solution bag on
the infusion stand.
4.2.4
Attach the arterial blood line
CAUTION!
Do not use the neonatal blood lines: A-5.128-B4, V-5.127-X, A-5.129-B4 or
V-5.129-X with AK 98 dialysis machine as these neonatal blood lines are not
intended for use with the AK 98 dialysis machine.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...