A:82
AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
HCEN12745 Revision 10.2016
Program version 2.xx
4)
Hold the segment in place and close the
blood pump door.
5)
Press lightly on the blood line to place it in
the upper guide.
6)
Remove protective caps and attach the
end of the arterial blood line firmly to the
dialyzer; press and turn simultaneously.
WARNING!
Check that the arterial blood
line is firmly attached to the
dialyzer to prevent patient
blood loss.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...