AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
607 Wrong disinfectant
607 Wrong disinfectant
Check disinfectant. To continue press
Confirm.
Appears:
When the dialysis machine detects a too high or too low
disinfectant concentration.
The appearance of the attention can be preset.
Machine actions:
None.
Operator action:
1.
Check that the disinfectant set in the dialysis machine is
the same as the actual disinfectant connected to the
machine.
2.
Press
Confirm
to repeat the fill-up phase.
Mute time:
Permanent.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...