AK 98 Dialysis Machine - Disinfection and cleaning
A:149
3)
Press
Auto schedule
.
4)
Select day.
5)
Press
Off
.
6)
Press
OK
.
10.3.8
Integrated heat disinfection with a WRO 300 H unit
10.3.8.1
Integrated heat disinfection with a WRO 300 H unit
An integrated heat disinfection program has an extra phase compared to normal heat
disinfection. In this phase hot water from the water purification unit WRO 300 H slowly
flows through the dialysis machine. This means that also the inlet water tube is
disinfected. Therefore, choose integrated heat disinfection if possible. For more
information, see Section 11
“Disinfection with the AK 98 dialysis machine and WRO
10.4
Chemical disinfection
10.4.1
About chemical disinfection
During the chemical disinfection program the machine is filled with a concentrated
disinfectant
. Concentrated disinfectant is mixed with water to the correct
concentration in the machine. The diluted solution fills all parts of the fluid path. After a
certain time (dwell time) the fluid path is rinsed and drained.
Heat disinfection can be replaced by chemical disinfection. But note that decalcification
(CLEANCART C cartridge or liquid citric acid) cannot be replaced by chemical
disinfection.
10.4.2
Start a chemical disinfection
Procedure
1)
Press the
Disinfection
button.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...