AK 98 Dialysis Machine - Technical data and specifications
A:191
NOTE!
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE!
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Recommended separation distances between portable and mobile RF communications
equipment and the AK 98 dialysis machine
The AK 98 dialysis machine is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the AK 98 dialysis
machine can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the AK 98
dialysis machine as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter(m)
150kHz - 80MHz
80MHz - 800MHz
800MHz - 2500MHz
0.01
0.04
0.04
0.07
0.1
0.12
0.12
0.23
1
0.35
0.35
0.7
10
1.2
1.2
2.3
100
3.5
3.5
7.0
Rated maximum
output power of
mobile phone
-
-
Bluetooth/GSM/CDMA
/LTE/RFID/WLAN
-
-
0.35
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
d
in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE!
At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE!
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
14.6.4
Expected service life
AK 98 dialysis machine
10 years
BPM cuff
2 years
Table 14-24. Expected service life
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...