HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Alarms
B:27
110 High diastolic blood pressure
110 High diastolic blood pressure
To clear the alarm press Confirm.
Appears:
When the patient's diastolic blood pressure is outside the
set high alarm limit.
Machine actions:
None.
Operator action:
1.
Check the blood pressure history:
- Press the
BPM
button.
- Select the
History
tab.
- Press
History
.
- Observe the measurement results from the treatment.
2.
Do standard patient checks and take measures for
patients with circulatory disturbances.
3.
If desired, measure the blood pressure manually (see
Section 8.3
“Direct blood pressure measuring”
on
page A:130 in Operators handbook).
4.
If possible, adjust the alarm limits to the patient’s status:
- Press the
BPM
button.
- Select the
Alarm
tab.
- Press
Set limits
.
- Press
Diastolic High
and adjust the alarm limit.
5.
Press
OK
.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...