HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Alarms
B:41
124 Technical error
124 Technical error
Contact technical service.
Appears:
When there is a technical fault in the machine that affects
the patient safety.
Machine actions:
The actions depend on the kind of technical error. The
machine sometimes automatically enters a state which
makes it impossible to continue the treatment.
Operator action:
1.
Press
Confirm
.
2.
Open the
Error list
:
a)
Press the
Blood
button.
b)
Select the
Tools
tab.
c)
Press
Service
.
d)
Select the
Error list
tab.
Read and note the error code text on the screen. If
more than one technical error have been generated, the
most recent is displayed at the top of the list.
3.
Technical error during treatment: call an authorised
technician. If necessary during treatment, manually
return the blood to the patient. See Section 3.6.3
“Return the blood to the patient manually”
on
page A:71 in Operators handbook.
WARNING!
A restart attempt after technical error during treatment
may create a potential hazard for the patient.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...