B:60
AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
217 TMP is too high
217 TMP is too high
Appears:
When the transmembrane pressure (TMP) has reached the
set high alarm limit.
Machine actions:
None.
Operator action:
1.
Press the
Fluid
button.
2.
Select the
TMP
tab.
3.
Check for a cause: unexpected patient weight loss or
issues with the needle, extracorporeal circuit or dialyzer
(clogged dialyzer), and take appropriate actions.
4.
Do standard patient checks, for example measure
blood pressure.
5.
If possible, increase the treatment time, decrease the
set UF volume or increase the blood flow rate.
6.
If possible, adjust the high TMP alarm limit.
7.
Confirm that the actual TMP value and alarm limits are
in accordance with the actual UF rate and UF
coefficient.
8.
If necessary, end the treatment and call an authorised
technician.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...