AK 98 Dialysis Machine - Maintenance handling
HCEN12745 Revision 10.2016
Program version 2.xx
13.1
Maintenance
The contents of this chapter concerns the maintenance of the AK 98 dialysis machine
that can be carried out by the operator of the machine.
Maintain the dialysis machine according to recommendations in AK 98 Operator’s
manual.
CAUTION!
To ensure proper functionality of the dialysis machine, only perform maintenance
activities that are described in AK 98 Operator’s manual.
CAUTION!
To ensure proper functionality of the dialysis machine, only an authorised service
technician should open the casing and do repairs and modifications.
CAUTION!
Do not use the dialysis machine if it is damaged in any way, or does not work as
described in this operator’s manual.
13.2
Blood Pump Rotor
13.2.1
Maintenence of the blood pump rotor
Clean the blood pump rotor when it is dirty.
13.2.2
Clean the blood pump rotor
Procedure
1)
To open the pump cover: pull the opening
tab towards you, and at the same time
press firmly at the middle of the cover.
2)
Turn the handle clockwise until the blood
pump rotor loosens.
Wipe the rotor and the interior of the pump
housing with a lint-free cloth, moistened
with ethanol (70 %) or isopropanol (60 %).
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...