HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
A:101
WARNING!
Check that the arterial blood
line is connected to the
rinseback solution to prevent
patient blood loss.
5)
Connect the arterial blood line to the rinse
back solution. Open the clamp on the rinse
back line and/or the arterial bloodline.
CAUTION!
Keep an eye on the venous
needle and venous pressure
during the rinse back, since
there is no automatic
supervision at this stage.
6)
Press the
Blood pump
button to start the
blood pump.
7)
When the rinse back is finished and the
blood pump stops, disconnect the venous
blood line from the access.
4.3.2
Confirm disconnect patient
WARNING!
To avoid air infusion, ensure that the patient is physically disconnected from the
arterial and venous bloodlines before confirming
Disconnect patient
. When
Disconnect patient
has been confirmed the air detector is deactivated. The
patient must not be reconnected.
Procedure
1)
Press
Disconnect patient
.
2)
Confirm that the arterial and venous blood lines are disconnected.
3)
Remove the venous blood line from the priming detector.
4.3.3
Machine aftercare
Before you begin
CAUTION!
To ensure that the patient is not connected when doing machine aftercare, check
that the arterial and venous blood lines have been completely disconnected from the
patient.
Procedure
1)
Turn the dialyzer so that the blue
connectors are up.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...