35
10. Retest Fire.
Move the X gantry to the top Y position. Fire another black dot onto the
thermal paper. You should now have two black dots (one from the upper Y position and
one from lower Y position). Move the X gantry up and down and verify that the red dot
tracks from one black dot to the other black dot.
11. Adjust Mirror #1.
Move the X gantry to the lowest Y position. The red dot should now
overlap with one of the black dots. Adjust mirror #1 until the red dot is in the middle of the
two black dots
.
12. Retestfire.
Use a fresh piece of thermal paper and fire a test dot with the X gantry still in
the lowest Y position. Move the X gantry to the top position and fire another test dot. Now
the two black dots should be closer together.
13. Adjust.
Move the X gantry back to the lowest Y position. Now adjust the red dot until it
moves to the center of the two black dots.
14. Repeat Until Aligned.
Repeat the process of firing two black dots on the top and on the
bottom Y positions and adjusting mirror #1 to move the red dot to the center of the two
black dots until the two black dots from the top and bottom Y positions are centered on top
of each other.