Detector III
Error message
Description
Cause of the fault / solution
signals
signals but cannot as
there is not enough
memory .
y ou hav e taken multiple measurements at this measuring
point. The characteristic v alues are alw ay s sav ed, but none
or only some of the time signals are sav ed.
Input
ov ermodulated!
The v ibration signal is
too large to be measured
by the Detector.
There are tw o possible causes for this:
The input signal is v ery large (>50 g). Use another
sensor (w ith a sensitiv ity of 10 mV/g, for ex ample).
The strength of the signal increased after the amplifier
w as adjusted due to an impact on the machine, for
ex ample. The v ibrations must remain more or less
constant throughout the entire measurement.
No RPM signal!
The rotational speed
signal is missing.
Possible causes:
The trigger sensor is not correctly aligned w ith the reflex
mark.
The reflex mark w as not glued on.
The trigger sensor is too close to the reflex mark (<10
cm). Increase the distance.
The trigger sensor has been incorrectly configured
in Trendline (incorrect supply v oltage, for ex ample).
You hav e not connected the trigger sensor to the AUX
port
of the Detector.
Rotational speed
unstable!
The rotational speed
signal is fluctuating.
Possible causes:
The trigger sensor is not correctly aligned w ith the reflex
mark.
The trigger sensor is too close to the reflex mark (should
be at least 10 cm).
Machine rotational speed fluctuates.
The rotational
speed not w ithin
range.
The acquired speed is
outside the rotational
speed band defined in .
Reduce the speed of the machine or use the current
rotational speed as the new nominal rotational speed.
No conf. for RFID No configuration found
for the RFID entry .
This error occurs w hen Trendline sends the RFID status for
a measuring point before a balancing, CM or run up/coast
dow n configuration w as sent. Make sure that there is a
configuration for the measuring point on the Detector. The
Trendline database may contain an error. Please contact
support@fis-serv ices.com.
Could not find at
least one
configuration
Detected RFID tags and
configured measuring
points do not match.
The Detector found at least one RFID tag w ithout a
measuring point configuration. The missing measuring point
is therefore not offered for selection for the measurement.
Multiple IDs read.
Please repeat
Tw o or more tags w ere
found w hile try ing to
assign an RFID tag.
The assignment is not possible as the Detector found at least
tw o tags in the RFID Reader read area. Remov e the
unnecessary RFID tags from the read area.