258
the Standard amplitude/phase measurement
process.
Multiple measurements
You can repeat a free measurement just like a measurement. For more
information on multiple measurements, see the section Multiple measurements
.
6.12 Single measurements
With the Detector, you can perform single measurements to the ISO 10816
characteristic value determination, for "listening" to the measurement point, for
temperature and rotational speed measurements as well as for the ICP sensor
test.
Press on
Single measurements
in the main menu.
Please note that single measurements cannot b e saved.
ISO 10816
This menu item allows you to perform a single CM measurement for recording
characteristic values as per ISO 10816 (class 1-4). Please note that you cannot
save this measurement!
The following fixed settings are used to perform for the ISO 10816 single
measurement:
Class 1
Class 2
Class 3
Class 4
Pre-alarm threshold
1.8 mm/s
2.8 mm/s
4.5 mm/s
7.1 mm/s
Main alarm threshold
4.5 mm/s
7.1 mm/s
11.2 mm/s
18.0 mm/s
Sensitiv ity of sensor [mV/g]
100
ISO 10816 classes
249
231