5-point linearity protocol
!
Note!
The specified measuring accuracy is a typical value
94, "Performance characteristics".
With the production of the 5-point linearity protocol the measuring system (FDU9x sensor and
FMU9x transmitter electronic) is adjusted exactly to one another and the measuring accuracy is
optimized for the specified range.
To realize this, the parameter "zero distance" is fine adjusted. After a reset the value for the zero
distance has to be re-parameterized in the service menu according to the data on the associated 5-
point linearity protocol for the FDU9x sensor. Please contact the Hauser service.
6.2
Configuration of the measurement
6.2.1
First setup
!
Note!
This chapter describes the commissioning of the Prosonic S via the display and operating module.
Commissioning via FieldCare is similar. For further instructions refer the FieldCare Online Help.
After switching on the power supply for the first time, the instrument asks for a number of operating
parameters:
!
Note!
By pressing
you can return to the previous parameter
(e.g. in order to correct the value).
All these parameters can also be changed at a later point of time in the "device properties/operating
parameters" and "device properties/language" parameter sets.
1.
Select the display language.
a. Press
or
to move the marking bar to the
desired language.
b. Press
to confirm your selection.
L00-FMU90xxx-07-00-00-yy-027
2.
Select the unit for distance measurements.
L00-FMU90xxx-07-00-00-en-028
3.
Select the temperature unit.
L00-FMU90xxx-07-00-00-en-029