Wiring
Prosonic S FMU95
26
E Hauser
4.8
Potential equalization
4.8.1
Potential equalization in the field housing
Example
L00-FMU95xxx-04-00-00-xx-009
1
The wire is already connected on delivery
#
Warning!
The grounding line of the sensors FDU91F/93/
95/96 and FDU83/84/85/86 must be
connected to the local potential equalization
system
after a maximum of 30 m (98 ft)
,
(
20). The metallic terminal block (1) in
the field housing can be used for this.
L00-FMU90xxx-04-00-00-xx-006
1
90 … 253 VAC
10,5 … 32 VDC
N
L
PE
FDU91F/93/95/96
(FDU83/84/85/86)
1
FMU95
Address
T
erm.
DP
of
f
on
of
f
on
SW
HW
1
2
3
4
5
6
7
8
1
2
3
4
A(N)
66
B(P)
65
Display
POWER
Sync
Fuse
40
39
Service
3
2
1
11
10
9
1
FDU-Sensor
YE
BK
RD
14
13
12
2
YE
BK
RD
17
16
15
3
YE
BK
RD
20
19
18
4
YE
BK
RD
23
22
21
5
YE
BK
RD
26
25
24
6
YE
BK
RD
29
28
27
7
YE
BK
RD
32
31
30
8
YE
BK
RD
35
34
33
9
YE
BK
RD
38
37
36
10
YE
BK
RD
FDU-Sensor
GNYE
FDU91F/93/95/96
(FDU83/84/85/86)
GNYE