Error codes
The error code consists of 6 digits with the following meaning:
• Digit 1: Type of error
– A: alarm
– W: warning
– E: error (the user can define if the error behaves like an alarm or a warning.)
• Digits 2 and 3:
indicate the input channel, output channel or the relay to which the error refers. "00" means that
the error does not refer to a specific channel or relay.
• Digits 4-6:
indicate the error according to the following table.
Example:
W 01 641
• W: Warning
• 01: sensor input 1
• 641: loss of echo
Code
Description of error
Remedy
A 00 100
software version does not fit to hard-
ware version
A 00 101
checksum error
full reset and recalibration required
A 00 102
checksum error
full reset and recalibration required
W 00 103
initializing - please wait
if the message does not disappear after a couple of seconds:
replace electronics
A 00 106
downloading - please wait
wait for completion of the download
A 00 110
checksum error
full reset and recalibration required
A 00 111
A 00 112
A 00 114
A 00 115
electronics defective
switch instrument off/on;
if the error persists:
call Hauser service
A 00 116
download error
repeat download
A 00 117
hardware not recognised after exchange
A 00 125
electronics defective
replace electronics
A 00 152
checksum error
full reset and recalibration required
W 00 153
initializing
if the message does not disappear after a couple of seconds:
replace electronics
A 00 155
electronics defective
replace electronics
A 00 164
electronics defective
replace electronics
A 00 171
electronics defective
replace electronics
A 00 180
synchronization faulty
check synchronization wiring (
A 00 183
hardware not supported
check if the installed board complies with the order code of
the instrument;
call Hauser service
A 01 231
...
A 10 231
sensor 01 - 10 defective - check con-
nection
check for correct connection of the sensor
(
A 01 281
...
A 10 281
temperature measurement 01 - 10
defective - check connection
check for correct connection of the sensor
(