Installation
Prosonic S FMU95
14
E Hauser
3.4.4
Adaption plate
If an opening of 138 x 138 mm and the remote display of the Prosonic FMU860/861/862 are
already present, you can use the adaption plate (Order Code: 52027441, (
It is inserted into the remote display of the FMU860/861/862.
!
Note!
The adapter plate is mounted directly in the housing of the old remote display of the FMU86x series.
The housing of the remote display of the FMU86x is the holder for the adapter plate and the new
remote display of the FMU90/FMU95 in the format 96 x 96 mm (3.78 x 3.78 in).
L00-FMU90xxx-06-00-00-xx-006
1
Remote display of the FMU90 with adaption plate
2
Opening of the remote display of the FMU860/861/862
3.5
Mounting of the sensors
Information on the mounting of the sensors can be found in the following documents:
• Technical Information TI00189F (for FDU8x)
• Technical Information TI00396F (for FDU9x)
These documents are supplied with the sensors.
3.6
Installation check
After installing the device, carry out the following checks:
• Is the device damaged (visual inspection)?
• Does the device correspond to the meausring point specifications such as process temperature,
process pressure, ambient temperature, measuring range etc?
• If available: Are the measuring point number and labelling correct?
• Is the instrument sufficiently protected against rainfall and direct sunlight?
• For the field housing: Are the cable glands tightened correctly?
• Is the instrument securely mounted to the DIN rail or the mounting help (visual inspection)?
• For the field housing: Are the screws of the terminal compartment cover securely tightened
(visual inspection)?