Troubleshooting
Prosonic S FMU95
76
E Hauser
W 01 501
...
W 10 501
no sensor selected for channel 01 - 10
allocate sensor
("level" or "flow" menu)
A 01 502
...
A 10 502
Sensor 01 - 10 not recognized
Enter type of sensor manually
("level" or "flow" menu, submenu "basic calibration".
A 00 511
no factory calibration present
A 01 512
...
A 10 512
mapping in process
wait for completion of mapping
W01 521
...
W10 521
new sensor 01 - 10 detected
W01 601
...
W10 601
non-monotonic linearisation curve for
level 01 - 10
re-enter linearisation
(s. "level" menu")
A 01 604
...
A 10 604
faulty calibration for level 01 - 10
adjust calibration
(s. "level" menu)
W01 611
...
W10 611
linearisation points level 01 - 10: num-
ber < 2
enter further linearisation points
(s. "level" menu)
E 01 641
...
E 10 641
no usable echo sensor 01 - 10
check basic calibration for the respective sensor
(s. "level" or "flow" menu)
A 01 651
...
A 10 651
Safety distane reached for sensor 01 -
10: danger of overfilling
Error disappears if the level is out of the safety distance again.
Possibly, the function "acknowledge alarm" must be used
(s. "safety settings" menu)
E 01 661
...
E 10 661
temperature sensor 01 - 10 too high
W01 691
...
W10 691
filling noise detected sensor 01 - 10
W00 801
simulation level switched on
switch off level simulation
(s. "level" menu)
W01 802
...
W10 802
simulation sensor 01 -10 switched on
switch off simulation
A 00 820
...
A 00 832
Different units for calculation of average
value or sum
Check the units of the respective basic calibrations
(s. "level" menu)
Code
Description of error
Remedy