Prosonic S FMU95
Operation
E Hauser
33
General key combinations
The following key combinations do not depend on the menu position:
5.2.2
The operating menu
Structure of the menu
The parameters of the Prosonic S are organized in an operating menu (consisting of a main menu
and several submenus). Parameters which are related to each other are comprised in a common
parameter set. To simplify the navigation within the menu, a five-digit position code is displayed
with each parameter set.
• The
first digit (1)
specifies the submenu
1)
:
–
L:
"level"
–
A:
"safety settings"
–
O:
"output/calculations"
–
D:
"device properties", "calibr. display" and "sensor management"
–
I:
"system information"
–
S:
"service" (only available if the service password has been entered)
Diagrams of the submenus can be found in the chapter "Operating menu".
Key combination
Meaning
Escape
• While editing a parameter: Exit the editing mode without accepting the changes.
• Within the navigation: Move upwards to the previous layer of the menu.
Increase contrast
Increases the contrast of the display module.
Decrease contrast
Decreases the contrast of the display module.
Locking
Locks the instrument against parameter changes.
The instrument can only be unlocked again by the keys.
LVL 1 appl.para
L1004
tank shape :dome ceiling
medium property:liquid
process cond:standard liq.
Identification of the parameter sets:
1
Submenu
2
Number of the associated input or output
3
Number of the parameter set within the submenus
1)
Depending on the instrument version, the installation environment and the selected operating mode, some of the submenus may not be present.
3
2
1
Содержание Prosonic S FMU95
Страница 2: ......
Страница 104: ...Appendix Prosonic S FMU95 104 Endress Hauser...
Страница 107: ...Prosonic S FMU95 Index Endress Hauser 107...
Страница 108: ...www endress com worldwide BA00344F 00 EN 13 12 71164423 FM SGML 9 0 71164423...