Prosonic S FMU95
Installation
E Hauser
13
3.4
Mounting the remote display and operating module
3.4.1
Scope of delivery
If the Prosonic S is ordered with the display for cabinet door mounting, the following is contained
in the scope of delivery:
• Display and operating module, 96 x 96 mm (3.78 x 3.78 in)
• 4 retainers (with nuts and screws)
• Connection cable (3 m (9.8 ft)) for connection to the transmitter
(preassembled with suitable plugs).
3.4.2
Dimensions of the separate display and operating module
L00-FMU90xxx-06-00-00-xx-004
3.4.3
Mounting
1.
Cut an opening of 92 x 92 mm (3.62 x 3.62 in) into the intended mounting position (e.g.
cabinet door).
2.
Insert the remote display module into the opening and fix the retainers as shown in the
following figure:
L00-FMU90xxx-17-00-00-xx-002
96
55
max. 6
92
min. 11
~
0,5 kg (1.10 lbs)
(0.43)
(3.62)
(0.24)
(3.78)
(2.17)
mm (in)
1.
3.
4.
2.
Содержание Prosonic S FMU95
Страница 2: ......
Страница 104: ...Appendix Prosonic S FMU95 104 Endress Hauser...
Страница 107: ...Prosonic S FMU95 Index Endress Hauser 107...
Страница 108: ...www endress com worldwide BA00344F 00 EN 13 12 71164423 FM SGML 9 0 71164423...