Hazardous Area
X-STREAM O
2
Transmitter
vi
Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
QUICK START GUIDE
FOR X-STREAM
SYSTEMS
1. Install the probe with transmitter electronics in an appropriate location
on the stack or duct. Refer to Section 2: Installation for additional
information on selecting a location for the Hazardous Area X-STREAM
O
2
Transmitter.
2. Connect reference air to the Hazardous Area X-STREAM O
2
Transmitter.
3. Connect HART communications, either through a handheld
communicator like the Rosemount 375 Field Communicator, or through
Emerson's AMS Suite PC software, or on an AMS-aware DCS.
NOTE
For a transmitter with optional Xi Enhanced Interface, the 375 Field
Communicator must be connected only to the 4 to 20 mA signal loop between
the Xi and the control room or data acquisition system. Connecting the 375
Field Communicator between the transmitter and Xi will cause communication
errors and affect system operation.
4. Apply AC power to the cell heater.Then connect the loop-power for the
transmitter electronics.
NOTE
Although the transmitter electronics are loop-powered, they are not
intrinsically safe.
5. Allow approximately 45 minutes for the cell heater to warm up to the
736°C (1357°F)
heater setpoint. The 4-20 mA signal will remain at a
default value of 3.5 mA throughout this warm-up period. Once warm, the
probe will be reading oxygen, and the 4-20 mA signal will be a reading
based on the default range of 0-10% O
2
.
6. Perform a manual calibration. Refer to "Quick Reference Guide Manual
Calibration Instructions" that follow. Proceed through all setup menus,
including calibration setup. Calibrate the probe with cal gas and record
the slope, impedance, and cell constant on a copy of the Calibration
Record shown on page 5-8 of this manual.
Содержание Rosemount X-STREAM O2 Transmitter
Страница 2: ......
Страница 4: ......
Страница 8: ...Hazardous Area X STREAM O2 Transmitter TOC 4 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 10: ...Hazardous Area X STREAM O2 Transmitter ii Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 32: ...Hazardous Area X STREAM O2 Transmitter 1 16 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 58: ...Hazardous Area X STREAM O2 Transmitter 2 26 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 66: ...Hazardous Area X STREAM O2 Transmitter 3 8 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...