Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
6-5
Hazardous Area
X-STREAM O
2
Transmitter
X-STREAM O
2
TRANSMITTER
PROBE REPAIR
Each of the procedures that follow details how to remove and replace a
specific component of the Hazardous Area X-STREAM O
2
Transmitter. Part
replacement kits are available for each of the transmitter components
replaced in this repair section. Refer to Section 7: Replacement Parts for kit
part numbers.
Sensor Core or Diffuser
Replacement
1. See Figure 6-1. Loosen and remove the lock clip (22) from the cover
lock slots.
2. Remove left housing cover (18) to access the core.
3. Remove three screws (19).
4. Disconnect the sensor connector plug from electronics board (8). Pull
on the connector lanyard as shown in Figure 6-2.
NOTE
A connector plug with lanyard replacement kit is available in case the sensor
connector plug is worn or damaged. Refer to Section 7: Replacement Parts
for the replacement kit part number.
Figure 6-2. Disconnecting
Connector Plug
It is recommended that the X-STREAM be removed from the stack for all service activities.
The unit should be allowed to cool and be taken to a clean work area. Failure to comply may
cause severe burns.
Disconnect and lock out power before working on any electrical components. There is
voltage up to 240 VAC.
Содержание Rosemount X-STREAM O2 Transmitter
Страница 2: ......
Страница 4: ......
Страница 8: ...Hazardous Area X STREAM O2 Transmitter TOC 4 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 10: ...Hazardous Area X STREAM O2 Transmitter ii Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 32: ...Hazardous Area X STREAM O2 Transmitter 1 16 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 58: ...Hazardous Area X STREAM O2 Transmitter 2 26 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 66: ...Hazardous Area X STREAM O2 Transmitter 3 8 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...