Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
2-15
Hazardous Area
X-STREAM O
2
Transmitter
Optional Xi Enhanced
Interface
1. Remove cover screws from the front cover of the Xi Enhanced Interface.
Swing down the front cover of the interface box.
2. Pull out the I/O board on the right-hand side of the card rack inside the
Xi Enhanced Interface. If your system is configured to operate two
transmitter probes there are two I/O interface boards.
3. See Figure 2-12. Connect the 4-20 mA signal wires at J4 of the I/O
board. Attach the supplied ferrite clamp over the 4-20 mA OUT wires
that extent past the shield.
NOTE
Installation of the ferrite clamp over the 4-20 mA OUT wires is required for
compliance with the European EMC Directive.
Figure 2-12. Signal Connections Xi Enhanced Interface
Figure 2-13. Alarm Indicator Relay Terminals