Hazardous Area
X-STREAM O
2
Transmitter
6-12
Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
Cell Installation
Use the following instructions to install a new O
2
sensing cell.
1. Apply anti-seize compound to flange of new cell, Figure 6-9.
2. Place new corrugated washer onto cell flange with the ridges pointing
up as shown in Figure 6-9. Fit is tight. Hold gasket in place with one
finger then "walk" the gasket in place using another finger. Use care not
to bend or kink the washer.
3. Apply anti-seize compound to exposed surface of corrugated washer.
4. Install the new cell (4, Figure 6-1) with corrugated washer (3) into the
end of sensor core assembly (1).
5. Apply anti-seize compound to threads of cell retainer (5) and mating
threads in heater housing.
6. Thread cell retainer (5) into the heater housing. Use cell removal tool
and 1/4 inch diameter pin to tighten cell retainer to 30 ft-lbs (40,67 N
·
m).
7. Tighten screw (24, Figure 6-5) to secure contact and thermocouple
assembly (5).
8. Remove the setscrews from diffuser (6, Figure 6-1).
9. Apply anti-seize compound to threads of setscrews and thread
setscrews three turns into diffuser.
10. Install diffuser (6) in the same orientation as the old diffuser. Install and
tighten the diffuser setscrews.
11. Apply a light coating of silicone grease onto the o-ring of sensor core
assembly (1).
12. Install sensor core assembly (1) in probe tube (7).
13. Install and tighten three screws (20) to secure sensor core assembly (1)
inside probe tube (7).
14. Connect the sensor core plug to electronics board (8).
15. Re-connect the cal gas and reference air hoses as shown in Figure 6-3.
The cal gas hose connects between the hose ports marked "C". The
reference air hose connects between the hose ports marked "R".
16. Recalibrate the probe per "O2 Calibration" in Section 4 of this instruction
manual.
If installing a ceramic or Hastelloy diffuser be sure to orient the rounded deflector so it points
upstream relative to the flue gas air flow.
Do not use excessive force when installing the reference air or cal gas hose onto the
connecting tube. Using excessive force can buckle the stainless steel tubing.
Failure to correctly install the hoses may allow gas leakage and result in erroneous O
2
readings. Be sure to correctly install the calibration gas and reference air hoses.
Содержание Rosemount X-STREAM O2 Transmitter
Страница 2: ......
Страница 4: ......
Страница 8: ...Hazardous Area X STREAM O2 Transmitter TOC 4 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 10: ...Hazardous Area X STREAM O2 Transmitter ii Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 32: ...Hazardous Area X STREAM O2 Transmitter 1 16 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 58: ...Hazardous Area X STREAM O2 Transmitter 2 26 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 66: ...Hazardous Area X STREAM O2 Transmitter 3 8 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...