Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
2-13
Hazardous Area
X-STREAM O
2
Transmitter
ELECTRICAL
INSTALLATION
All wiring must conform to local and national codes. Multiple wiring diagrams
are shown in this section. Always refer to the diagrams that apply to your
transmitter configuration and disregard all other wiring diagrams.
X-STREAM Transmitter
NOTE
To maintain proper earth grounding ensure a positive connection exists
between the transmitter housing and earth. The connecting ground wire must
be 14 AWG minimum. Refer to Figure 2-11.
NOTE
Line voltage, signal, and relay wiring must be rated for at least 105°C
(221°F).
Disconnect and lock out power before connecting the power supply.
Install all protective covers and safety ground leads after installation. Failure to install covers
and ground leads could result in serious injury or death.
To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation
of this equipment, connection to the main electrical power supply must be made through a
circuit breaker (min 10A) which will disconnect all current-carrying conductors during a fault
situation. This circuit breaker should also include a mechanically operated isolating switch.
If not, then another external means of disconnecting the supply from the equipment should
be located close by. Circuit breakers or switches must comply with a recognized standard
such as IEC 947.
The Xi Enhanced Interface can be installed in general purpose areas only. Do not install the
Xi in hazardous areas or in the vicinity of flammable liquids.
To maintain explosion-proof protection of the X-STREAM O
2
Transmitter , all cable entry
devices and blanking elements for unused apertures must be certified flameproof, suitable
for the conditions of use and be properly installed.
Содержание Rosemount X-STREAM O2 Transmitter
Страница 2: ......
Страница 4: ......
Страница 8: ...Hazardous Area X STREAM O2 Transmitter TOC 4 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 10: ...Hazardous Area X STREAM O2 Transmitter ii Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 32: ...Hazardous Area X STREAM O2 Transmitter 1 16 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 58: ...Hazardous Area X STREAM O2 Transmitter 2 26 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 66: ...Hazardous Area X STREAM O2 Transmitter 3 8 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...