Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
6-27
Hazardous Area
X-STREAM O
2
Transmitter
Figure 6-22. Power Supply
Board Wiring
6. Hold the AC input wiring to the right and slide the Power Supply board
out of the Xi enclosure.
7. Install the new mounting bracket (9, Figure 6-14) on the new Power
Supply board (10, Figure 6-14). A new bracket and mounting screws are
provided in the replacement kit.
8. Slide the Power Supply board into the mating slots in the Xi Enclosure.
Make sure the board is correctly aligned in the slots.
9. Install and tighten the bracket mounting screws. Two new screws are
provided in the replacement kit and should be used if the existing
screws are damaged.
10. Connect the ribbon cable to the Power Supply board. A new ribbon
cable (7, Figure 6-14) is provided in the replacement kit and should be
used if the existing cable is damaged.
11. Connect the AC power plug to the Power Supply board.
12. Swing the Xi cover up and tighten the four mounting screws.
Xi Front Panel
Replacement
Use the procedure that follows to replace the front panel on the Xi Enhanced
Interface. Replacement kits with and without the CPU board are available.
Use the instructions that apply to the replacement kit you have.
Disconnect and lock out power before working on any electrical components.
Содержание Rosemount X-STREAM O2 Transmitter
Страница 2: ......
Страница 4: ......
Страница 8: ...Hazardous Area X STREAM O2 Transmitter TOC 4 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 10: ...Hazardous Area X STREAM O2 Transmitter ii Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 32: ...Hazardous Area X STREAM O2 Transmitter 1 16 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 58: ...Hazardous Area X STREAM O2 Transmitter 2 26 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 66: ...Hazardous Area X STREAM O2 Transmitter 3 8 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...