Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
Hazardous Area
X-STREAM O
2
Transmitter
http://www.raihome.com
Section 6
Maintenance and Service
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-1
Maintenance Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-2
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-2
Automatic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-2
Manual Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-2
Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-3
Removal and Installation of Probe . . . . . . . . . . . . . . . . . . page 6-3
X-STREAM O2 Transmitter Probe Repair . . . . . . . . . . . . . page 6-5
Sensor Core or Diffuser Replacement . . . . . . . . . . . . . . . page 6-5
Cell Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-8
Replace Contact Pad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-8
Cell Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-12
Replace Contact & Thermocouple Assembly . . . . . . . . . page 6-13
Replacing Gas/Air Hoses, O-rings and Spacers . . . . . . . page 6-14
Electronics Board Replacement . . . . . . . . . . . . . . . . . . . . page 6-14
Blind Cover(s) Replacement . . . . . . . . . . . . . . . . . . . . . . . page 6-15
Xi Components Replacement . . . . . . . . . . . . . . . . . . . . . . page 6-18
I/O Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-19
AC Relay Board Replacement . . . . . . . . . . . . . . . . . . . . . . page 6-24
Power Supply Board Replacement . . . . . . . . . . . . . . . . . . page 6-26
Power Supply Board Wiring . . . . . . . . . . . . . . . . . . . . . . . page 6-27
Xi Front Panel Replacement . . . . . . . . . . . . . . . . . . . . . . . page 6-27
DR Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-30
OVERVIEW
This section identifies the calibration methods available and provides the
procedures to maintain and service the Hazardous Area X-STREAM O
2
Transmitter.
Install all protective equipment covers and safety ground leads after equipment repair or
service. Failure to install covers and ground leads could result in serious injury or death.
Содержание Rosemount X-STREAM O2 Transmitter
Страница 2: ......
Страница 4: ......
Страница 8: ...Hazardous Area X STREAM O2 Transmitter TOC 4 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 10: ...Hazardous Area X STREAM O2 Transmitter ii Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 32: ...Hazardous Area X STREAM O2 Transmitter 1 16 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 58: ...Hazardous Area X STREAM O2 Transmitter 2 26 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 66: ...Hazardous Area X STREAM O2 Transmitter 3 8 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...