Hazardous Area
X-STREAM O
2
Transmitter
6-22
Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
Figure 6-18. /O Board Flame Safety Interlock Wiring
11. Reinstall wiring for Alarm Outputs, Flame Status Input and/or SPS/IMPS
as applicable. See Figure 6-17 and Figure 6-18 for wiring diagrams.
12. Connect the ribbon cable to the I/O Board. A new cable is supplied and
should be used if the old one is damaged.
13. Swing the Xi cover up in place and tighten the four screws.
14. Prior to operating the X-STREAM and the Xi Enhanced Interface, all
optional software enhancements (previously enabled or not) must be
enabled. Notify Rosemount Analytical Inc. and reference the following
part numbers to enable the related software options:
NOTE
For enhanced software upgrades or to enable optional software features
previously used in your Hazardous Area X-STREAM O
2
Transmitter
configuration, contact Rosemount Analytical Inc. at 1-800-433-6076.
15. Recalibrate the O
2
Transmitter according to the applicable calibration
instructions in Section 4: Startup and Operation.
Part Number
Software Option
6A00269G01
Enhanced Software Option Upgrade, Stoichiometric Function
6A00269G02
Enhanced Software Option Upgrade, Programmable Reference Function
6A00269G03
Enhanced Software Option Upgrade, 850
°
C Process Function
Содержание Rosemount X-STREAM O2 Transmitter
Страница 2: ......
Страница 4: ......
Страница 8: ...Hazardous Area X STREAM O2 Transmitter TOC 4 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 10: ...Hazardous Area X STREAM O2 Transmitter ii Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 32: ...Hazardous Area X STREAM O2 Transmitter 1 16 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 58: ...Hazardous Area X STREAM O2 Transmitter 2 26 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 66: ...Hazardous Area X STREAM O2 Transmitter 3 8 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...