Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
6-3
Hazardous Area
X-STREAM O
2
Transmitter
Replacement Parts
Refer to Section 7: Replacement Parts for individual replacement parts and
part replacement kits. Part replacement kits are available for each of the
components discussed in this section.
REMOVAL AND
INSTALLATION
OF PROBE
Remove Probe
1. Turn off power to the system. Lock out and tag out power to the
X-STREAM probe.
2. Shut off instrument air and calibration gases at the cylinders.
3. Disconnect calibration gas and instrument air lines from the probe.
4. See Figure 6-1. Loosen and remove the lock clip (22) from the cover
lock slots.
5. Remove right housing cover (12).
6. Disconnect all signal and power wiring to the probe.
7. Remove insulation at probe head to access the mounting bolts or
threaded adapter.
8. Unbolt or unscrew the X-STREAM probe assembly from the stack and
take it to a clean work area.
9. Allow the X-STREAM probe to cool to a comfortable working
temperature.
Install Probe
1. Bolt or thread the X-STREAM probe assembly to the stack. Use
anti-seize compound or Teflon tape on mounting threads.
2. Install the insulation at the probe head.
3. Connect all signal and power leads at the probe. Refer to Section 2:
Installation, for detailed wiring instructions.
4. Connect the calibration gas and instrument air lines to the probe.
5. Install right housing cover (12, Figure 6-1).
6. Insert the lock clip (22) into the cover lock slots and tighten screw (24).
7. Turn on calibration gas and instrument air.
8. Restore power to the system; refer to Section 4: Startup and Operation.
When the probe is at operating temperature, calibrate the probe per "O2
Calibration".
NOTE
Recalibration is required whenever an electronic board or sensing cell is
replaced.
Failure to use anti-seize compound or Teflon tape on threads could result in thread galling,
probe damage, and excessive difficulty in removal of probe.
Содержание Rosemount X-STREAM O2 Transmitter
Страница 2: ......
Страница 4: ......
Страница 8: ...Hazardous Area X STREAM O2 Transmitter TOC 4 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 10: ...Hazardous Area X STREAM O2 Transmitter ii Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 32: ...Hazardous Area X STREAM O2 Transmitter 1 16 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 58: ...Hazardous Area X STREAM O2 Transmitter 2 26 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 66: ...Hazardous Area X STREAM O2 Transmitter 3 8 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...