Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
4-7
Hazardous Area
X-STREAM O
2
Transmitter
SYSTEM PARAMETER
DESCRIPTIONS
Among the parameters available through the Xi and 375 Field Communicator
menus are a number of "System Parameters". The system parameters define
variables that configure the Xi in the transmitter system. System parameters
are described in the following table.
CPU
PARAMETER
NAME
UNITS
PARAMETER DESCRIPTION
Y
Version
--
Software version number for the CPU board.
Y
Build Num
--
Software build number for the CPU board.
Y
Comm Status
--
Communication status between the CPU and I/O boards.
Y
Restart Cntr
--
Software restarts counter for the CPU board.
N
Poll Addr
--
Polling Address - Address used to identify a Field Device; changeable by the user to control.
N
Serial Number
--
I/O board serial number.
N
Features
--
Available advanced software features for the I/O board.
ON: feature supported.
OFF: feature not supported.
N
EE Erase Count
--
This is for nonvolatile memory diagnostic only.
Y
Num Probe
--
Number of probes configuration:
If set to 1, use One Probe configuration for the Main display.
If set to 2, use Two Probes configuration for the Main display.
Y
Probe Sel
--
This parameter is applicable only if the Num Probe configuration is set to 2. It defines which
probe is to be displayed.
Y
Probe 1 Enable
--
Probe 1 state. If enabled, the probe menu will be displayed otherwise it will not be displayed.
(No, Yes)
Y
Probe 2 Enable
--
Probe 2 state. If enabled, the probe menu will be displayed otherwise it will not be displayed.
(No, Yes)
Y
LineX Probe
--
Probe number for line x. (1, 2)
Y
Line 1 Data
--
Main display, line 1 data: 0 displays PV (O
2
), 1 displays SV (O
2
Temp)
Y
Line 2 Data
--
Main display, line 1 data: 0 displays PV (O
2
), 1 displays SV (O
2
Temp)
Y
Line 3 DataL
--
Main display, line 3 left hand side data:
0 displays PV (O
2
)
1 displays SV (O
2
Temp)
2 displays TV (Cell Imp)
3 displays 4V (Cell mV)
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
Y
Line 3 DataR
--
Main display, line 3 right hand side data:
0 displays PV (O
2
)
1 displays SV (O
2
Temp)
2 displays TV (Cell Imp)
3 displays 4V (Cell mV)
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
Y
Line 4 DataL
--
Main display, line 4 left hand side data:
0 displays PV (O
2
)
1 displays SV (O
2
Temp)
2 displays TV (Cell Imp)
3 displays 4V (Cell mV)
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
Y
Line 4 DataR
--
Main display, line 4 right hand side data:
0 displays PV (O
2
)
1 displays SV (O
2
Temp)
2 displays TV (Cell Imp)
3 displays 4V (Cell mV)
4 displays CJC Temp
5 displays AO
6 displays OP Mode
7 displays Tag
Y
Revert Time
min
Xi display reverts to main display time. Also locks screen if password protection is enabled.
Y
Language
--
0 displays English
1 displays Spanish
2 displays German
Y
Contrast
--
Display contrast: (120 to 200)
Y
Alarms
--
See section 5, Troubleshooting, ”Alarm Indications”.
Y
Password
--
Security Password
Y
Enable Password
--
Enables/disables security password protection
Содержание Rosemount X-STREAM O2 Transmitter
Страница 2: ......
Страница 4: ......
Страница 8: ...Hazardous Area X STREAM O2 Transmitter TOC 4 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 10: ...Hazardous Area X STREAM O2 Transmitter ii Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 32: ...Hazardous Area X STREAM O2 Transmitter 1 16 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 58: ...Hazardous Area X STREAM O2 Transmitter 2 26 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 66: ...Hazardous Area X STREAM O2 Transmitter 3 8 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...