Instruction Manual
IM-106-910C, Rev. 1.0
November 2009
4-9
Hazardous Area
X-STREAM O
2
Transmitter
N
Y
Prev Cal Imp
Ohm
Previous Cell Impedance. This is the sensor resistance that was calculated as a result of
previous successful calibration. There are ten calibration results. Index 1 is the most recent and
Index 10 is the least recent sensor resistance measured.
N
Y
Failed Slope
mV/Dec Failed calibration slope.
N
Y
Failed Const
mV
Failed calibration constant.
Y
Y
Cal Result
mV
Calibration result.
N
Y
Delta Imp
--
Delta impedance since last calibration.
N
Y
Cal Step
--
This represents the step of the calibration cycle is in.
N
Y
Time Remain
sec
Time remaining in the present calibration cycle state.
Y
Y
O2 Slope
mV/Dec O2 slope. This is the slope value that will be used to calculate O2.
Y
Y
O2 Const
mV
O2 constant. This is the constant value that will be used to calculate O2.
N
Y
T90 Filter
sec
Analog output T90 time. It represents the time to take a step change in oxygen to reach 90% of
the final value at the filter output.
N
Y
O2 Cell Ref
%
O2 sensor reference gas percentage. It allows using a sensor reference gas other than air.
N
Y
O2 Temp SP
degC
O2 sensor temperature set point. It allows measurement of oxygen with an elevated sensor
temperature. (0=736 degC set point; 1=834 degC set point)
N
Y
Features
--
Advanced software features.
(0=Stoichiometer; 1=Programmable Reference; 2=Elevated Temperature)
Y
Y
O2 URV
%
Primary variable (O2%) upper range value.
Y
Y
O2 LRV
%
Primary variable (O2%) lower range value.
N
Y
O2 AO Range
--
Analog output polarity. (0=4-20 mA; 1=20-4 mA)
Y
Y
O2 Alarm Level
--
O2 alarm level. (0=3.5 mA; 1=21.1 mA)
N
Y
Alarm Relay 1
--
Alarm Relay 1 mode.
(no alarm; unit alarm; low O2 alarm; low O2/unit alarm;
Cal recommended; Cal recommended/unit alarm; low O2/Cal recommended;
low O2/unit alarm/Cal recommended)
N
Y
Alarm Relay 2
--
Alarm Relay 2 mode. (no alarm; unit alarm; low O2 alarm; low O2/unit alarm
Cal recommended; Cal recommended/unit alarm; low O2/Cal recommended;
low O2/unit alarm/Cal recommended; Heater relay)
N
Y
Low O2 Alm
%
Low O2 alarm threshold.
N
Y
Hi Temp Alm
%
High temperature alarm threshold.
N
Y
AO Tracks
--
Analog output track O2 sensor measurement during a calibration. (No, Yes).
N
Y
Auto Cal
--
Enable/disable automatic calibration.
No = Set to Manual calibration mode.
Yes = Set to Automatic calibration mode.
N
Y
Start On CalRec
--
Start automatic calibration on Cal Recommended state. (No, Yes)
Y
Y
Cal Gas 1
%
Test Gas 1 value. This is the actual value of the gas being applied during the Test Gas 1 phase
of a calibration.
Y
Y
Cal Gas 2
%
Test Gas 2 value. This is the actual value of the gas being applied during the Test Gas 2 phase
of a calibration.
Y
Y
Gas Time
sec
Test Gas application time. This is the length of time test gases are applied to the O2 probe
during low or high Test Gas phase of a calibration.
Y
Y
Purge Time
sec
Test Gas purge time. This is the length of time before the output will be returned to the process
reading after a calibration.
N
Y
Cal Interval
hr
Automatic calibration interval. The number 9999 disables the automatic timed calibration.
N
Y
Next Cal Time
hr
Time remaining until the next automatic periodic calibration. The number 9999 disables the next
automatic timed calibration.
Y
Y
T EE Val
--
Transmitter board nonvolatile memory diagnostic.
N
Y
IO EE Val
--
I/O board nonvolatile memory diagnostic.
Y
Y
Version
--
Software version number for the Transmitter.
N
Y
Version
--
Software version number for the I/O board.
Y
Y
T Restart Cntr
--
Software restarts count for the Transmitter.
N
Y
IO Restart Cntr
--
Software restarts count for the IO board.
Y
Y
Alarms
--
Current Alarms (See section 5, Troubleshooting, ”Alarm Indications”.)
TX
I/O
PARAMETER
NAME
UNITS
PARAMETER DESCRIPTION
Содержание Rosemount X-STREAM O2 Transmitter
Страница 2: ......
Страница 4: ......
Страница 8: ...Hazardous Area X STREAM O2 Transmitter TOC 4 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 10: ...Hazardous Area X STREAM O2 Transmitter ii Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 32: ...Hazardous Area X STREAM O2 Transmitter 1 16 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 58: ...Hazardous Area X STREAM O2 Transmitter 2 26 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...
Страница 66: ...Hazardous Area X STREAM O2 Transmitter 3 8 Instruction Manual IM 106 910C Rev 1 0 November 2009 ...