LEVEL 3:
n
Reverse on load - motor A
(safety feature)
The device automatically reverses both
motors when the patient’s resistance
(load) exceeds the set value.
Adjustable levels for reverse on load
feature: 1 - 25. At level 1, very low resist-
ance will cause the device to reverse; at
level 25, a high resistance is required to
initiate the reversal.
Default: level 25
n
Reverse on load - motor B
(safety setting)
The device automatically reverses both
motors when the patient’s resistance
(load) exceeds the set value.
Adjustable levels for reverse on load
feature: 1 - 25.
At level 1, very low resistance will cause
the device to reverse; at level 25, a high
resistance is required to initiate the
reversal.
Default: level 25
m
Caution!
Patient hazard —
The reverse on load feature is a safety
measure to protect the patient in the
event of cramps, spasms, locked joints
and similar situations. The manufac-
turer cannot be held liable for misuse of
this feature.
n
Motor A ON/OFF
To permit fully isolated movements, the
motors can individually be turned on
and of. Motor A efects the extension
and lexion movements, while motor b
efects the pronation and supination
movements.
For an isolated pronation/supination
movement, set motor A to the appropri-
ate position (extension/lexion), then
turn the motor of.
in normal mode the display indicates the
symbol “
OFF
” during therapy, instead
of the programmed extension/lexion
angles.
Default: motor A ON
n
Motor B ON/OFF
To permit fully isolated movements, the
motors can individually be turned on
and of. Motor A efects the extension
and lexion movements, while motor
b efects the pronation and supination
movements.
For an isolated extension/lexion move-
ment, set motor b to the appropriate
position (pronation/supination), then
turn the motor of.
in normal mode the display indicates the
symbol “
OFF
” during therapy, instead of
the programmed pronation/supination
angles.
Default: motor B ON
Note!
Please note that one motor (A or b)
must always be turned on. Otherwise
the following symbol will be displayed
when you press
START
:
LEVEL 4:
n
Isolation protocol
For this special function, both motors
are turned on, but they will never move
simultaneously.
These are the steps of the function:
• To begin with, motor A moves to the
Follow these steps to program pauses:
• Access menu level 2, then press the
parameter key to select the special
function
.
• The display shows the symbol for
the special function and the selected
pause durations.
The display shows the symbol for
the special function and the selected
pause durations.
top: lexion/supination pause (
).
bottom: extension/pronation pause
(
).
The lexion/supination pause duration
appears in a box.
• To change the lexion/supination
pause, press the “+” or “-” key.
select a duration of 10 seconds, for
instance.
• Press the
parameter key
again when
done. The frame automatically moves
to the line below, indicating that the
extension/pronation pause
can
be adjusted.
The
parameter key
is only used to
toggle between these two functions.
• To modify the value, press the “+”
(plus) or “-” (minus) key.
select a duration of 5 seconds, for
instance.
• save the programmed values by
pressing the
STOP
key, then press the
START
key to initiate the treatment.
n
Therapy timer
Default setting
is
continuous operation
.
A clock symbol in the upper right-hand
corner of the display identiies the
con-
tinuous mode of operation
. The clock
indicates the elapsed therapy time.
in the continuous mode, the device must
be stopped with the
STOP
key.
however, the therapy timer can be set in
steps of 1 minute to any value between
1 and 59 minutes and in steps of 30
minutes to any value between 1 and
24 hours.
when the time has elapsed,
the device switches automatically of
and stops in the middle between the set
limit values.
in this case, a circle replaces the clock
symbol. The circle ills as the therapy
time progresses.
n
Speed
The speed can be adjusted between 1 %
and 100 % in steps of 1 %.
1 % is equivalent to 14°/minute
100 % is equivalent to 230°/minute
Default: 100 %
n
Warm-up protocol
during warm-up, the patient will slowly
become used to the set maximum val-
ues for extension/lexion and pronation/
supination.
The warm-up protocol starts in the
middle between the two maximum
values set for extension/lexion and
pronation/supination. The range of
motion increases with each cycle, until
the programmed maximum values are
reached after a total of 15 cycles.
After reaching the maximum values, the
device automatically enters the normal
mode.
if a therapy time is activated, the system
restarts with the warm-up protocol
when the therapy time has elapsed and
the devices is started again.
during warm-up, the display will show
the symbol
in the upper, left-
hand corner.
Default: disabled