Section 7: Native Mode Commands
In-Sight
®
1700 Series Wafer Reader
72
SETTING
DESCRIPTION
<LightMode> Determines
either
a Bright Field or a Dark Field illumination effect on the mark.
0 = All lights are disabled
1 – 3 = The mark will appear dark on a light background
4 – 11 = The mark will appear light on a dark background
<LightPower>
The exposure time of the wafer reader’s CCD imager, in milliseconds.
0.000-33.000 (default = 2.000)
<Mark>
The type of wafer mark the selected Config will read.
0 = 2D Sym, T7 Data Matrix
1 = BC, BC 412
2 = BC, IBM 412
3 = Internal Use Only
4 = Chars, SEMI
5 = Chars, IBM
6 = Chars, Triple
<MarkColor>
Specifies whether Color settings will be tried during the tuning process.
0 = Disable Color tuning
1 = Enable Color tuning (default)
<NumFailed>
The number of failed reads for a Config.
0 – 99,999
<NumPassed>
The number of passed reads for a Config.
0 – 99,999
<NumTuneFailed>
The number of failed reads during a tune for a Config.
0 – 99,999
<NumTunePassed>
The number of passed reads during a tune for a Config.
0 – 99,999
<Optimization>
The index of the Optimization type for the specified Config, when ‘Chars, SEMI’ is the
selected Mark.
0 = Speed (default)
1 = Reliability
<Order>
The order in which enabled Configs will be tried until a successful read can be obtained.
0 = Last Best (default)
1 = In Order (0-9)
<Orientation>
The index of the Img Orientation setting for the specified Config.
0 = Normal (default)
1 = Mirrored horizontally
2 = Flipped vertically
3 = Rotated 180 degrees
<Passed>
The pass/fail status of the specified Config for the most recent read.
0 – 99,999
<Phi> (Region)
The curve of the Config's region.
-360 to 360
<ReadLog>
A record of the most recent read that passed.
0-9
<Region>
The region where the wafer mark is expected to appear in the field of view.
Row:
640 x 480 image resolution: 0 – 479
1024 x 768 image resolution: 0 – 767
High:
640 x 480 image resolution: 4 – 240 (default = 100)
1024 x 768 image resolution: 4 – 240 (default = 100)
Col:
640 x 480 image resolution: 0 – 639
1024 x 768 image resolution: 0 – 1023
Wide:
640 x 480 image resolution: 4 – 640 (default = 500)
1024 x 768 image resolution: 4 – 1024 (default = 600)
Theta: -179 to 180
Phi: -360 to 360
Содержание In-Sight 1721
Страница 1: ...IN SIGHT 1700 SERIES WAFER READER USER MANUAL...
Страница 2: ......
Страница 8: ...In Sight 1700 Series Wafer Reader vi...
Страница 58: ...Section 4 Managing Job Files In Sight 1700 Series Wafer Reader 48...
Страница 130: ...Section 8 Specifications In Sight 1700 Series Wafer Reader 120...
Страница 136: ...Updating the Wafer Reader s Firmware In Sight 1700 Series Wafer Reader 126...
Страница 140: ...Configuring Microsoft Windows Network Settings In Sight 1700 Series Wafer Reader 130...
Страница 154: ...In Sight 1700 Wafer Reader Mount Conversion In Sight 1700 Series Wafer Reader 144...
Страница 156: ...In Sight 1700 Series Wafer Reader www cognex com products InSight default asp Rev 16 May 2006...