Section 7: Native Mode Commands
In-Sight
®
1700 Series Wafer Reader
68
7.2.2 Put PortNum
Put PortNum(<Input>)
Specifies the port that will be used by the In-Sight wafer reader for transferring data or
images over the network.
NOTES
■
The wafer reader must be Offline when using the Portnum command.
■
The PortNum command must be executed before another device can connect to the specified port.
Inputs
:
0 to 65535
to specify the port number.
NOTE
A valid Port assignment is any unused number between 0 and 65535, excluding 21, 23, 68, 80,
502, 1069, 1070, 1212, 2222, 44818 and 50000 (reserved for In-Sight communications).
Output
: Status Code
The status codes are:
STATUS CODE
DESCRIPTION
1
The command was executed successfully.
0 Unrecognized
command.
-1
The port number is invalid.
-2
The command could not be executed, the wafer reader is Online, or the specified port is
already in use.
Sample Input and Output
PUT PORTNUM 3001
1
Содержание In-Sight 1721
Страница 1: ...IN SIGHT 1700 SERIES WAFER READER USER MANUAL...
Страница 2: ......
Страница 8: ...In Sight 1700 Series Wafer Reader vi...
Страница 58: ...Section 4 Managing Job Files In Sight 1700 Series Wafer Reader 48...
Страница 130: ...Section 8 Specifications In Sight 1700 Series Wafer Reader 120...
Страница 136: ...Updating the Wafer Reader s Firmware In Sight 1700 Series Wafer Reader 126...
Страница 140: ...Configuring Microsoft Windows Network Settings In Sight 1700 Series Wafer Reader 130...
Страница 154: ...In Sight 1700 Wafer Reader Mount Conversion In Sight 1700 Series Wafer Reader 144...
Страница 156: ...In Sight 1700 Series Wafer Reader www cognex com products InSight default asp Rev 16 May 2006...