MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
92
7.4. Communication / Networking
The TEC 3000 is equipped with two independent RJ-45 serial ports. These ports are intended for
connection to another MVE controller, PC, serial printer, or other RS-485 device. Up to 100 TEC
3000 controllers can be successfully networked.
NOTE: The TEC 3000 should never be connected directly to a LAN or
public telecommunications network.
Figure 12:
TEC 3000 serial ports and RS-485 4-wire pin diagram
Figure 13:
Standard RJ-45 cable assembly wiring details (PN 10740053). Wire colors may vary.
Table 6:
RS-485 Interface Specifications
COM Setup
bits/sec
Parity
Data bits
Stop bits
9600 N81
9600
None
8
1
9600 N82
9600
None
8
2
19200 N81
19200
None
8
1
19200 N82
19200
None
8
2
COM Type
ASCII
Printer
OFAF
MODBUS
Disabled
Transmission
Mode
Terminator
4-wire
Asynchronous
CR, LF
SERIAL
SERIAL
1
2
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...