MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
65
5. Press “ENTER”
The display will read “Timed Fill”. Use the “▲/▼” keys to
adjust and press “ENTER” to save.
6. Press “SETUP”
The display will read “Timed Fill Frequency”. Use the
“▲/▼” keys to adjust and press “ENTER” to save. It can
be set from 1 to 28 days.
7. Press “SETUP”
The display will read “Timed Fill Start Hour”. Use the
“▲/▼” keys to adjust and press “ENTER” to save.
8. Press “SETUP”
The display will read “Timed Fill Start Minute” Use the
“▲/▼” keys to adjust and press “ENTER” to save.
Timed Fill
DISABLED
Use ▲▼ to adjust
Press ENTER to save
Timed Fill Frequency
5 Days
Use ▲▼ to adjust
Press ENTER to save
Timed
Fill Start
Hour
20:00
Use ▲▼ to adjust
Press ENTER to save
Timed
Fill Start
Minute
20:00
Use ▲▼ to adjust
Press ENTER to save
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...