MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
154
Level Commands
BPFIL?
Bypass Status Query
Input: BPFIL?
Returned:
0 or 1
Comments:
Used to query the current hot gas bypass status where 0 = not
bypassing and 1 = bypassing.
FILAS?
Fill Time Alarm Status Query
Input: FILAS?
Returned:
0 or 1
Comments:
Used to query the Fill Time Alarm Status where 0 = off and 1 = on.
FILL
Set Fill Status
Input: FILL 0 or FILL 1
Returned:
No response
Comments:
Used to set the current Fill Status where 0 = Stop Fill and 1 = Start
Fill
FILL?
Fill Status Query
Input: FILL?
Returned:
0 or 1
Comments:
Used to query the current Fill Status where 0 = not filling and 1 =
filling.
FILLM?
Auto Fill Status Query
Input: FILLM?
Returned:
0 or 1
Comments:
Used to query the Auto Fill Status where 0 = disabled and 1 =
enabled.
FILT
Set Max Fill Time
Input: FILT XXX
Returned:
No response
Comments:
Used to set the maximum fill time in minutes from 30 to 240
minutes.
FILT?
Max Fill Time Query
Input: FILT?
Returned:
XXX
Comments:
Used to query the maximum fill time. Value returned in minutes.
FILTIM?
Fill Duration Query
Input: FILTIM?
Returned:
XXX
Comments:
Used to query the duration of the current fill. Value returned in
seconds.
HFIL
Set High Level Setpoint
Input: HFIL XXX.X
Returned:
No response
Comments:
Used to set the High Level Setpoint. Input value uses the current
level units, either inches or millimeters. There must be at least a 0.5 inch interval
between each level setting and alarm.
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...