MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
73
TEC3000 Default Settings
Parameter
Default Setting
High Temp Alarm
-110.0 °C / -166.0 °F / 163.2 K
Low Temp Alarm
-200.0 °C / -327.6 °F / 73.4 K
High Level Alarm
8.0 in / 205 mm
High Level Setpoint
7.0 in / 180 mm
Low Level Setpoint
5.0 in / 125 mm
Low Level Alarm
4.0 in / 100 mm
Defined 100%
8.0 in / 205 mm
Defined 0%
4.0 in / 100 mm
Level Offset
+1.3 in / +35 mm
Auto Fill
Enabled
Maximum Fill Time
60 minutes
Temperature Display Units
°C
Liquid Level Display Units
Inches
Hot Gas Bypass MENU
DISABLED
Hot Gas Bypass Temp Setpoint
-70 °C
Hot Gas Bypass Alarm Time Delay
5 minutes
COM 1 Type
ASCII
COM 2 Type
Disabled
Event Log Interval
240 minutes
GLOBAL PASSWORD
3 4 5 6