TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
87
10. Press “ENTER”
The display will read “Level Calibration Complete”.
11) Verify that the home screen reads the liquid level value that was just entered. In some
cases the controller should be restarted. Contact Technical support with any questions.
Level Calibration
Complete
Содержание MVE 1842P-150