MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
28
•
Connect the clear vinyl tubing to the TEC 3000 Level Input hose barb
and connect the other end
of the tubing to the 3-way purge valve
.
•
Ensure all of the plumbing assembly connections to the TEC 3000 wire harness are secure.
NOTE: Do not connect the battery backup (if equipped) to the wire harness until
later in the installation procedure.
•
Connect the 12-pin wire harness to the TEC 3000 wire harness connection.
•
If the freezer is equipped with battery backup measure the voltage at the end connector for
approximately 24VDC to 27VDC; if no voltage is present, the included battery fuse must be
installed before connecting the battery to the main wire harness. Open the battery enclosure and
unscrew the fuse harness. Install the fuse; close the fuse harness and the battery enclosure.
(This is done at the factory but should be verified at installation)
NOTE: Do not connect the battery backup to the main wire harness until the power supply has
been verified to power up the controller
•
Plug in the power supply to an appropriate wall outlet with the proper AC voltage. Avoid wall
outlets that are connected to emergency generator power if possible. Although an uninterruptible
power supply (UPS) is ideal to ensure continuous power, a surge protector or power conditioner
is recommended.
•
Plug the power supply into the TEC 3000 30 VDC power input. The TEC 3000 display should
illuminate and begin the startup sequence.
To avoid risk of electrical shock, this equipment must only be connected to a properly
grounded power source or outlet.
•
Following the startup sequence, the TEC 3000 may start to alarm. This is normal.
•
Press “Alarm Mute” to silence the audible buzzer for 30 minutes. For installation and startup
purposes, the alarm buzzer can be disabled; however, be sure to enable it when installation is
complete. For instructions, see the Alarm Buzzer page in Section 6 of this manual.
•
Connect the battery backup to the main wiring harness. While running on outlet power, the TEC
3000 supplies a steady 27 VDC trickle charge to the battery backup. The battery backup may
need to be charged for several hours before it is able to power the TEC 3000.
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...