MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
159
Sanitizing and Decontaminating Chart Aluminum and Stainless
Steel Products
MVE aluminum dewars are constructed with an aluminum inner, which utilizes a fiberglass neck support.
The stainless units are constructed with an inner entirely fabricated from stainless steel sheets. Any
cleaning solution that does not react with aluminum or stainless can be used in the sanitation process of
these dewars. In most cases, any household detergent or mild soap solution is suitable. The U.S. Custom
Service uses a solution called EXPOR for incoming shipments from abroad. This is mixed 9 parts water
mixed with sodium chloride & lactic acid. As mentioned above, however, any household cleaning solution
can be used. These include bleach, detergents, and mild soaps. Other cleaners and disinfectants that can
be safely used include hydrogen peroxide, chlorine/water and denatured alcohol.
NOTE: DO NOT USE
ANY PETROLEUM BASED CLEANING SOLUTION
. It is important that the inner vessel is thoroughly
rinsed with water and all cleaner residues have been removed. Spraying the solution into the inner vessel
is preferred, although agitation of the solution inside the inner will suffice. Vapor shippers and Doble units
will require filling the inner to its full capacity with cleaning mixture and then rinsing. Allow the unit to dry
thoroughly before putting into service. With vapor shippers, we suggest setting dewar inverted to drain
and dry. The process is not intended for use in older vapor shipper models manufactured prior to 1994.
The generally accepted practice of using 10% chlorine bleach with 90% water solution still holds as the
best method for decontamination. However, with some of the bovine and swine virus strains showing up
today, it is the conclusion of the agricultural professors at the University of Minnesota and Texas A & M
that an increased mixture of chlorine bleach to 30% and 70% water will kill most known viruses except
BSE.
To perform this sanitizing procedure, cover all inner surfaces with the solution, let stand for 30 minutes
and remove. Rinse the decontaminated surfaces with clean water and remove rinse water. Allow dewar to
dry before putting into service. For vapor shippers and Doble units, this means to place dewar on end
(inverted) and allow drying. Note: Vapor dewars can be used immediately after rinsing but may take
longer to recharge to 100% capacity.
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...