MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
100
7.4.3.
Printer Setup
The TEC 3000 compatible printer kit allows users to print a hard copy of the freezer’s status at
a set interval and as events occur. The TEC 3000 will print the current status of the freezer in
the event log format. The default print interval is 30 minutes.
NOTE: Adjusting the print interval will not affect the event log interval. These two
parameters are independent. The event log interval can be adjusted using ASCII
commands. All printed events are not necessarily logged in the event log; however, all
the events in the event log are printed.
TEC 3000 Printer Kit (P/N 11544943)
•
Epson Printer
•
RS-485 Converter and Adapters
•
DB-25 Cable
•
User’s Guide
Installation
1.
Setup printer as described in the included user’s guide
2. Assemble and connect RS-485 converter and adapters as shown below
3. Connect the printer to TEC 3000 serial port 1 or 2 via the RS-485 converter assembly
4.
Set the corresponding COM Setup to “9600 N81”
5. Set the corresponding COM Type to
“Printer”
6. Adjust the Print Interval to the desired value
7. Test setup by forcing an event or printing a new header or event
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...