MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
156
LFIL?
Low Level Setpoint Query
Input: LFIL?
Returned:
XXX.X
Comments:
Used to query the Low Level Setpoint. Value returned in either
inches or millimeters depending on the current level units. There must be at least
a 0.5 inch interval between each level setting and alarm.
LOLA
Set Low Level Alarm
Input: LOLA XXX.X
Returned:
No response
Comments:
Used to set the Low Level Alarm value. Input value uses the
current level units, either inches or millimeters. There must be at least a 0.5 inch
interval between each level setting and alarm.
LOLA?
Low Level Alarm Query
Input: LOLA?
Returned:
XXX.X
Comments:
Used to query the Low Level Alarm value. Value returned in either
inches or millimeters. There must be at least a 0.5 inch interval between each
level setting and alarm.
LOLS?
Low Level Alarm Status Query
Input: LOLS?
Returned:
0 or 1
Comments:
Used to query the Low Level Alarm Status where 0 = off and 1 =
on.
LOLM
Set Low Level Alarm Mask
Input: LOLM 0 or LOLM 1
Returned:
No response
Comments:
Used to enable or disable the Low Level Alarm where 0 = disabled
and 1 = enabled. Default setting is 1 = enabled.
LOLM?
Low Level Alarm Mask Query
Input: LOLM?
Returned:
0 or 1
Comments:
Used to query the Low Level Alarm Mask status where 0 = disabled
and 1 = enabled. Default setting is 1 = enabled.
LUNI
Set Level Units
Input: LUNI E, LUNI M, or LUNI %
Returned:
No response
Comments:
Used to set the LN2 level units where E = English (inches), M =
Metric (millimeters), and % = Percent.
LUNI?
Level Units Query
Input: LUNI?
Returned:
E, M, or %
Comments:
Used to query the current LN2 level units E = English (inches), M =
Metric (millimeters), and % = Percent.
RATE?
Current Liquid Usage Query
Input: RATE?
Returned:
XXX.X
Comments:
Used to query the current LN2 liquid usage rate.
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...