13289499 Rev H •
101
-----
Ω
(-)
(+)
7.5. Remote Alarm Tests
7.5.1.
Global Remote
The Global Remote contacts can be checked for continuity using a digital multimeter or ohm
meter.
•
Normal State (No Alarms)
There should be continuity between the COM and NC terminals
The COM
– NO circuit should be open
•
Alarm State
There should be continuity between the COM and NO terminals
The COM
– NC circuit should be open
7.5.2.
Discrete Contacts
The Discrete Contacts can be tested for continuity using a digital multimeter with diode check.
Since the Discrete Contacts are open collectors, the diode check function (
) should be
used.
•
Normal State (No Alarms)
All Discrete Contact terminals should be open with respect to COMMON
•
Alarm State
There should be continuity between the specific active alarm terminal and
COMMON. Inactive alarms should remain open with respect to COMMON
-----
Ω
(-)
(+)
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...