MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
91
7.3.3.2. Hot Gas Sensor Two Point Calibration
Wait for Bypass
Temp to stabilize
then press ENTER
Bypass Temp -195.8 °C
Remove Bypass Probe
from LN2 and press
ENTER
Wait for Bypass
Temp to stabilize
then press ENTER
Bypass Temp 0.0 °C
Bypass Probe two
Point calibration
complete
Higher User Level
Required use ▲▼ to
ENTER Password
0000
Press ENTER for
Add-on menus
or press SETUP for
next menu
Battery Status
On AC Power
Press ENTER to
Hot Gas Bypass menus
or press SETUP for
next menu
Hot Gas Bypass
ENABLED
Use ▲▼ to adjust
Press ENTER to save
Press ENTER for
Bypass Probe
Calibration or press
SETUP for next menu
Calibration Type
TWO POINT
Use ▲▼ to adjust
Press ENTER for next
Bypass Calibration
Place Bypass Probe
in LN2 and press
ENTER
Press ENTER for
Temperature Menus
or press SETUP for
next menu
Wait while probe
warms to room
temperature
180 seconds
Place Bypass Probe
in ice water and
press ENTER
1. Press “SETUP”
Controller will prompt for a password. A flashing cursor
will make it clear which digit is being changed. Use the
“▲/▼” keys to scroll to the appropriate number. Press
“ENTER” to advance the cursor to the next position.
2. Press “ENTER”
The display will
read “Temperature Menus”.
3. Press “SETUP”
Until the display reads “Add-on Menus”.
4. Press “ENTER”
The display will read “Battery Status”.
5. Press “SETUP”
Until the display reads “Hot Gas Bypass Menus”.
6. Press “ENTER”
The display will read “Hot Gas Bypass”.
7. Press “SETUP”
Until the display reads “Bypass Probe Calibration”.
8. Press “ENTER”
The display will read “Calibration Type”. Use the “▲/▼”
keys to set as “TWO POINT”.
9. Press “ENTER”
The display will read “Bypass Calibration”. Submerge
the sensing end of the bypass probe in LN2.
10. Press “ENTER”
The display will read “Wait for Bypass Temp to
stabilize”. Wait for the temp reading to stabilize while the
probe is submerged in LN2.
11. Press “ENTER”
The display will read “Remove Probe A from LN2”.
Remove the bypass sensor from the LN2.
10. Press “ENTER”
The display will read “Wait while probe warms to room
temperature”. Wait while the bypass sensor warms to
room temperature and the controller counts down.
11. Wait 180 seconds
After the controller counts down, the display will read
“Place Probe A in ice water”. Completely submerge the
sensing end of the probe in the ice water bath.
12. Press “ENTER”
The display will read “Wait for Temp A to stabilize”. Wait
for the displayed Temp A reading to stabilize while the
probe is submerged in the ice bath.
13. Press “ENTER”
The display will read “Bypass Probe two point calibration
complete”.
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...